beautypg.com

KD Scientific Legato 200 Series User Manual

Page 64

background image

64

Operating the KDS Legato 200 Series

K

D

S

L

e

g

a

to

2

0

0

S

e

ri

e

s

U

s

e

r’s

M
a

n

u

a

l

phone

508.429.6809

fax

508.893.0160

e-mail

[email protected]

web

www.kdscientific.com

KD Scientific Legato 200 Series Syringe Pump User’s Manual

Publication 5617-002-Rev F

Time commands

CITIME

Clears the infused time. Configuration Run Screen

mode only.

Command format:

citime

CTIME

Clears both the infused and withdrawn times.

Configuration Run Screen mode only.

Command format:

ctime

CTTIME

Clears the target time. Configuration Run Screen

mode only.

Command format:

cttime

CWTIME

Clears the withdrawn time. Configuration Run Screen

mode only.

Command format:

cwtime

ITIME

Displays the infused time. Configuration Run Screen

mode only.

Command format:

itime

Query response:

[##:]# seconds

[##]

or:

[##:]##:##:##

[##]

TTIME

Sets or displays the target time. Configuration Run Screen

mode only.

Command format:

ttime [{target time}]

Query response:

[##:]Target time not set

[##]

or:

[##:]# seconds

[##]

or:

[##:]##:##:##

[##]

WTIME

Displays the withdrawn time. Configuration Run Screen

mode only.

Command format:

wtime

Query response:

[##:]# seconds

[##]

or:

[##:]##:##:##

[##]