KD Scientific Legato 200 Series Syringe Pump User’s Manual
Publication 5617-002-Rev F
WRATE
Sets or displays the withdraw rate. Configuration Run
Screen mode only. The rate argument may be replaced
by “max” or “min” to set the maximum or minimum rate,
respectively. “lim” may be used to display the range
limits.
Command format:
wrate [max|min|lim|{rate} {rate units}]
Query response:
or:
Volume commands
CIVOLUME
Clears the infused volume. Configuration Run Screen
mode only.
Command format:
civolume
CTVOLUME
Clears the target volume. Configuration Run Screen mode
only.
Command format:
ctvolume
CVOLUME
Clears both the infused and withdrawn volumes.
Configuration Run Screen mode only.
Command format:
cvolume
CWVOLUME
Clears the withdrawn volume. Configuration Run Screen
mode only.
Command format:
cwvolume
IVOLUME
Displays the infused volume. Configuration Run Screen
mode only.
Command format:
ivolume
Query response:
SVOLUME
Sets or displays syringe volume. Configuration Run
Screen mode only.
Command format:
svolume
Query response:
or:
TVOLUME
Sets or displays the target volume. Configuration Run
Screen mode only.
Command format:
tvolume [{target volume} {volume units}]
Query response:
or:
WVOLUME
Displays the withdrawn volume. Configuration Run
Screen mode only.
Command format:
wvolume
Query response:
✐
NOTE: For commands requiring rate units,
use the following format: m, u, n, p / h, m, s
Ex: m/m = milliliter / minute
Ex: n/s = nanoliter / second