KD Scientific Legato 200 Series Syringe Pump User’s Manual
Publication 5617-002-Rev F
STOP / STP
Stops the pump.
Command formats:
stop
stp
WRUN
Runs the pump in the withdraw direction. Configuration
Run Screen mode only.
Command format:
wrun
Rate commands
CRATE
Displays the current rate that the motor is running at. A
valid response is returned only in dynamic situations
(while the pump is running). Configuration Run Screen
mode only.
Command format:
crate
Query response:
or:
DIAMETER
Sets or displays the syringe diameter in mm.
Configuration Run Screen mode only.
Command format:
diameter [{syringe diameter}]
Query response:
IRAMP
Sets or displays the infusion rates while ramping.
Configuration Run Screen mode only.
Command format:
iramp [{start rate} {start units} {end rate} {end units}
{ramp time in seconds}]
Query response:
or:
IRATE
Sets or displays the infusion rate. Configuration Run
Screen mode only. The rate argument may be replaced
by “max” or “min” to set the maximum or minimum rate,
respectively. “lim” may be used to display the range
limits.
Command format:
irate [max|min|lim|{rate} {rate units}]
Query response:
or:
WRAMP
Sets or displays the withdraw rates while ramping.
Configuration Run Screen mode only.
Command format:
wramp [{start rate} {end rate} {rate units}
{ramp time in seconds}]
Query response:
or:
✐
NOTE: To clear a ramp, use CTTIME
command.