KD Scientific Legato 200 Series Syringe Pump User’s Manual
Publication 5617-002-Rev F
5. Screw down the syringe clamp knob (6) until the
syringe barrel clamp (10) is tight against the
syringe barrel. Do not overtighten, especially
when using glass syringes.
6. Secure the syringe flanges by pushing the retaining
brackets (4,5) tightly against the syringe flanges
using the bracket clamping knobs (7,8). The
syringe plunger should be secured by the pusher
block retaining bracket and the flange of the
syringe should be secured by the syringe holder
block retaining bracket.
7.
Tighten the mechanical stop collars (9) on the
pump to prevent excess pusher block travel.
✐
NOTE: The syringe barrel clamp (10)
may need to be flipped depending on your
syringe size. To do so, completely unscrew
and remove the syringe barrel clamp knob
from the syringe block. Then remove lock
nut (11). Flip the bracket and reassemble.
SYRINGE LOADING
2
5
4
1
3
9
7
8
10
6
> 30 cc syringes
phone
508.429.6809
fax
508.893.0160
web
www.kdscientific.com
✐
NOTE: For infuse/withdraw models make sure
the plunger is secured in the retaining bracket
and the barrel of the syringe is secured in the
retaining bracket.
The Standard Legato 200 Series can accommodate up to two
syringes. Any type of syringe, glass, plastic or stainless steel
can be used. The pump can accommodate from 0.5 ul to 140
ml syringes. The syringe barrel clamp should be flipped for
syringes larger than 30 ml. For syringes smaller than 30 ml,
the syringe clamp should be used as shown.
1.
Squeeze the release tabs (1) and move the pusher
block (2) to the approximate length of the syringe.
2. Unscrew the bracket clamping knobs (7,8) to
loosen the retaining brackets (4,5) on the pusher
block (2) and the Syringe holder block (3).
3. Unscrew the syringe barrel clamp by unscrewing
the syringe barrel clamp knob (6).
4. Place the syringe on the syringe holder block (3).
Ensure the barrel flange is with in the syringe holder’s
block’s retaining bracket (5) and the plunger flange is
in the pusher’ blocks retaining bracket.
< 30 cc syringes
11
!
CAUTION: Be sure mechanical stop
collars are positioned properly to
prevent the plunger from bottoming
out on glass syringes.