I/O EVENT TRIGGERS

The KDS Legato 200 Series pump permits easy

connection to external I/O devices via a variety of

interfaces, including a built-in 15-pin D-sub connector.

The KDS Legato 200 Series software incorporates simple

commands for incorporating I/O controls into your pump

Programs. Integrated commands for control of an

external valve, multi-pump synchronization, and general

trigger signals are integrated in the software. The

following procedure describes the use of the Advanced

Options I/O commands. Following the procedure is a

description of the selection menu options for each of the

I/O command types. Details for the electrical connection

controls and pin-out description for the 15-pin User I/O

Connector are provided in Appendix D of this manual.

The KDS Legato 200 Series manages trigger events

through specific pins on the I/O Connector port on the

back of the pump. The following details define pin logic:

Triggers are communicated via Pin 4 or Pin 5.

Valve control signals are communicated via Pin 8.

Pump synchronization commands are

communicated via Pin 6.

Event signals are communicated via Pin 2.