KD Scientific Legato 200 Series User Manual
Page 74
74
Appendices
K
D
S
L
e
g
a
to
2
0
0
S
e
ri
e
s
U
s
e
r’s
M
a
n
u
a
l
KD Scientific Legato 200 Series Syringe Pump User’s Manual
Publication 5617-002-Rev F
Syringe Loading - 4 x 140 & 6/10 Multi-Racks
1.
Raise or remove Clamp Bar (6) by unscrewing
thumb screw (9).
2. Loosen T-Knobs (7), located under the Syringe
Holder (2), and slide the Retaining Bracket (3)
on the Syringe Holder open.
3. Adjust the Pusher Block to the approximate length
of the syringe(s) by squeezing the Release Tabs (1)
and sliding the Pusher Block into position.
4. Loosen the three thumb screws on the Anti-Syphon
Bar (4) to provide enough room for the plunger
flange of the syringe(s). If using the 6/10 rack, loosen
the two thumb screws on the outside of the Plunger
Holder (8). Flip the Plunger Holder open.
5. Place the syringe(s) on the Syringe Holder (2)
ensuring that the barrel flanges are within the
Syringe Holder’s Retaining Bracket (3) and the
plunger flanges are between the Pusher Bar (5)
and Anti-Syphon Bar (4).
6. Push the Syringe Holder’s Retaining Bracket (3)
tight against the barrel flanges and tighten into
place using the two T-Knobs (7).
7.
Tighten the Anti-Syphon Bar (4) using the three thumb
screws. If using the 6/10 rack, flip the Plunger
Holder back into place (8) and secure by
tightening thumb screws.
8. Lower or replace the Clamp Bar (6) and tighten
thumb screw until Clamp Bar is firm against syringe(s).
✐
NOTE: If using fewer syringes than the
maximum number of syringes, try to place
them symmetrically about the lead screw to
present a balanced load.
4 x 140 Multi-Rack
6/10 Multi-Rack
7
5
4
8
1
2
9
6
3
1.
Release Tabs
2.
Syringe Holder
3.
Retaining Bracket (Syringe Holder)
4.
Anti-Syphon Bar
5.
Pusher Bar
6.
Clamp Bar
7.
T-Knobs
8.
Plunger Holder
9.
Thumbscrew
phone
508.429.6809
fax
508.893.0160
web
www.kdscientific.com