11 Alarm messages

43

2014-06-12

11 Alarm

messages

The following alarm messages can be shown in alternation with the temperature display:

Alarm display

Cause

Remedy

9999 flashes

Gone above measured value
The measured value is too large, is
outside the range, or a probe has
broken.

h

Check probe and connecting cable for damage or
short-circuit

v

Chapter 4.2 “Connection diagram”

h

Check that the correct probe has been set or
connected

v

Chapter 7.1 “C111 Analog inputs”

Gone below measured value
The measured value is too small, is
outside the range, or a probe short-
circuit has occurred.