JULABO SC 2500w Professional Series Recirculating Coolers User Manual
Page 70

Remote control
70
Command
Parameter
Response of recirculating cooler
in_par_11
none
Tn control parameter of the cascade controller.
in_par_12
none
Tv control parameter of the cascade controller.
in_par_13
none
Adjusted maximum internal temperature of the cascade
controller.
in_par_14
none
Adjusted minimum internal temperature of the cascade
controller.
in_par_15
none
Band limit (upper)
in_par_16
none
Band limit (lower)
in_mode_01 none
Selected
setpoint:
0 = SETP 1
1 = SETP 2
2 = SETP 3
3 = Last setpoint setting was carried out
through an external programmer
in_mode_02 none
Selftuning
type:
0 = Selftuning „OFF“
1 = Selftuning „ONCE“
2 = Selftuning „ALWAYS“
in_mode_03
none
Type of the external programmer input:
0 = Voltage 0 V to 10 V
1 = Current 0 mA to 20 mA
in_mode_04 none
Internal/external temperature control:
0 = Temperature control with internal sensor.
1 = Temperature control with external Pt100 sensor.
in_mode_05 none
Recirculating cooler in Stop/Start condition:
0 = Stop
1 = Start
in_mode_08 none
Adjusted control dynamics
0 = aperiodic
1 = standard
in_hil_00
none
Max. cooling power (%).
in_hil_01
none
Max. heating power (%).