Recirculating Cooler

61

SP = Setpoint Programmer

 Press

.

The temperature value adjusted and set on the external voltage or
current source is displayed in line 2 of the VFD COMFORT-
DISPLAY for control purposes.

This EPROG input enables the use of different voltage and
current values as program parameters.

20

40

60

80

100

200

300

250

150

°C

1 2 3 4 5 6 7 8 9 10 V

2 4 6 8 10

14

18 20 mA

A

B


UW – Setting the LOW value:

: (See below

)

1.

Adjust and set the lowest desired value on the voltage or
current source resp. (Example A: 1 V).
Wait appr. 30 seconds.

2.

Assign a lower temperature threshold value to this adjusted
voltage/current value by pressing the appropriate buttons on
the keypad of the instrument (Example A: 20

C ) and set by

pressing enter

.

OW – Setting the HIGH value:

(See below

)

1.

Adjust and set the highest desired value on the voltage or
current source resp. (Example A: 10 V).
Wait appr. 30 seconds.

2.

Assign an upper temperature threshold value to this
adjusted voltage/current value by pressing the appropriate
buttons on the keypad of the instrument (Example A: 200

C)

and set by pressing enter

.

XXXXX

UW °C

20.00

XXXXX

OW °C

200.00

Example B in the diagram serves to illustrate that the end
point values are freely selectable (Ex: 8 mA and 16 mA).



XXXXX

SP152.00

I xx.xx

Example out of diagram A:

 Adjusting the voltage source for an output of 7.6 V!

The instrument calculates this value from the rise angle of the two
predecided end points (in example A: 7.6 V correspond to an external
setpoint temperature of 152.0

C ).

After returning the VFD display to standard display by pressing

,

this value is displayed in line 2 (Example: SP 152.00

C).

Notice:

If this adjustment is not correctly performed at two different points, the setpoint setting
will be incorrect.

Important:
The usable temperature range between > UW < and > OW < is limited to the
configured working temperature range of the circulator resp. of the unit combination
(working temperature range see technical specifications on page 11)