Menu functions

50


The started programmer


XXXXX

S XX.X
STEP X

A

S XX.X
XXhXXxx

B

S XX.X
XXhxx

C1

S XX.X
I xxx.x

C2

S XX.X
E xxx.x

D1

S XX.X
RUN

D2

S XX.X
PAUSE


The started programmer indicates the actually calculated setpoint S XX.X in
line 2. This value increases within the time period >Px TIM< until the target
temperature >Px SEP< of the section is reached.
If time in a section is set to „0“, the next section starts only as soon as the
target temperature is reached.
Use the edit keys

to scroll in line 3. The display changes in intervals

of approx. 4 seconds between the valid section (STEP X) and

A Remaining time of the valid section

or

B Remaining total time: Profile x number of repetitions or
C Actual bath temperature

I xxx.x – internal actual value or
E xxx.x – external actual value

D RUN – the programmer is started or

PAUSE – the profile was interrupted with the key

. Time is stopped

and the temperature is maintained constant at the last calculated
setpoint value.

Continue: Press

Interrupting / Termination of a profile

A profile can be terminated by pressing

.

Power failure with the programmer started:
– The reaction of the circulator is the same as when switched off and on
again with the mains switch. The VFD COMFORT-DISPLAY indicates
"OFF".
– If the AUTOSTART function is activated, the programmer starts again at
a point approx. 20 seconds before the interruption took place. However,
an uncontrolled change of the bath temperature happened.


Setting time/date for the start

XXXXX

TIME
14h25ss

XXXXX

DATE
19/1203
MMJJ

>TIME<

Enter the start time. Example: 14:25 hrs

>DATE<

Enter the start date. Example: 19 December 2003

Check the setting of the internal real time clock (see page 44).


Status at the end of the profile

XXXXX

PS END
SETP

>PS END<

(STBY / SETP)

Set the status for the end of the profile.
With the parameter STandBY the circulator enters the –OFF– state.
With the parameter SETPoint the circulator maintains the temperature at
the value of the last section.