Offmode pmp off – JULABO SC 2500w Professional Series Recirculating Coolers User Manual

Page 42

background image

Menu functions

42

9.3.3.

OFF-MODE – Pump motor on / off

XXXXX

OFFMODE
PMP OFF

Factory setting:
PMP

OFF

Normally the circulating pump is switched via the start/stop signal.
However, if circulation should be maintained also for the -OFF- condition,
the parameter >PMP ON< needs to be set.

Possible parameters:

PMP ON

Pump motor on

PMP OFF

Pump motor off

1.

Press the

button until the submenu > OFFMOD < is displayed.

2.

Select the parameter with the keys

.

(PMP ON / PMP OFF)

3.

Press enter

to store the selected parameter.

Continue: Press

/

or

In case of an alarm state, a shutdown of the pump motor is still
effected.







9.3.4.

RESET – Factory settings

Factory setting:
NO

Use this to reset all values to factory setting (except date and time).

A RESET can only be carried out in the -OFF- condition.

Possible parameters:

NO / YES

1.

Press the

button until the submenu > RESET < is displayed.

2.

Select the parameter with the keys

(NO / YES).

3.

Press enter

to store the selected parameter.

As long as the message -RUN- appears all parameters a reset to
factory settings.