8.2. Over and Sub temperature warning functions

Over temperature

Sub temperature

XXXXX

SUBTMP
-94.9



If for a sensitive temperature application task adherence to a working
temperature value > SETP < is to be supervised, then set over and sub
temperature warning values.
In the example below, the > SETP < of 85 °C is surrounded by the values
> OVERTEMP < 87 °C and > SUBTEMP < 83 °C. The electronics
immediately registers when the actual temperature attains a temperature out
of the limits and it follows a reaction according to what is set in the menu
item >LIMITSR<.
(see chapter 8.2.1. Change-over of the warning function to shutdown
function)

OVERTMP
87.0

SUBTMP
83,0

XXXXX

SETP
85.0

1.

Press the

button until the menu item > OVERTMP < or

>SUBTMP< is displayed.

2.

Set value:

Use the cursor keys to move left or right on the display until the
numeral you wish to change blinks.

Use the increase/decrease arrows

to change the selected

numeral (-, 0, 1, 2, 3, ... 9).

3.

Press enter

to store the value.

The warning functions are only triggered when the actual bath
temperature, after start from the „OFF“ or „rOFF“ mode, lies within the
set limits for 3 seconds.

Recommendation:

Set the high temperature warning value > OVERTMP < at 5 °C to 10 °C
above the working temperature setpoint.

Set the low temperature warning value > SUBTMP < at 5 °C to 10 °C below
the working temperature setpoint.