Menu program – integrated programmer, Menu program – JULABO SC 2500w Professional Series Recirculating Coolers User Manual
Page 48
Menu functions
48
25
50
10 20 30 40 50 60 10 20 30 40 50 60 10 [min]
75
100
150
200
°C
t
external system
Internal bath
BAND HIGH
BAND LOW
Sense of a band limit:
Ensures gentle heating for the
objects requiring temperature
application.
Protects glass reactors, for
example, from thermal shock.
The INTMAX: and INTMIN: values are
superordinated to this band limit.
9.6. MENU PROGRAM – Integrated programmer
XXXXX
MENU
PROGRAM
1 Profil
10 Sections (STEP)
99 Repetitions
PROGRAM: Press enter
to switch to menu level 2
>PS STEP<
Program start at section ....
> PS RUNS <
Number of profile repetitions 1 ... 99
> PS GO <
Start type (NOW/TIME)
> P TIME <
Start time (hh:mm)
>P DATE<
Start date (TT/MM.JJ)
>PS END<
Status at the end of the profile (STBY/SETP)
(Standby or last setpoint)
> PE STEP<
Edit profile (1 ... 10)
>Px SEP <
Target temperature of section ...
>Px TIM<
Time period of section ...
>Px DEL<
Delete a section (STEP) (YES/NO)
> PE DEL<
Delete a profile (YES/NO)
The integrated programmer allows any desired temperature program
sequences to be realized. Such a temperature sequence is called profile. A
profile consists of individual sections defined by duration (t:) and target
temperature. Target temperature is the setpoint (T:), that is achieved at the
end of a section. The programmer uses time and temperature difference
values within a section to calculate a temperature ramp.