Operator responsibility – Safety instructions
12
Professional series
Temperature selection
digital
via keypad
indication on VFD COMFORT-DISPLAY
remote control via personal computer
indication on monitor
Temperature indication
°C
VFD COMFORT-DISPLAY
Resolution
°C
0.1
Temperature control
ICC - Intelligent Cascade Control
Electrical connections:
Computer interface RS232
External Pt100 sensor
Optional
Programmer input
-100 °C to 400 °C = 0 - 10 V or 0 - 20 mA or 4 - 20 mA
Input for the signal of a flow meter or external manipulated variable
Temperature recorder outputs
0 - 10 V (0 V = -100 °C, 10 V = 400 °C)
0 - 20 mA (0 mA = -100 °C, 20 mA = 400 °C)
4 - 20 mA (4 mA = -100 °C, 20 mA = 400 °C)
Standby input
for external emergency switch-off
Alarm output
for external alarm signal
Table 1
Pump capacity
Bath fluid: Water, Silicone -oil
XX.XX
PUMP
1
XX.XX
PUMP
2
Circulating pump:
P3 P3
Flow rate max.
Lpm at 0 bar
30 33
Pressure max.
bar at 0 liter
1.8 3.5
Bath fluid:
Galden
®
eg.
Fluorinert
®
3283
Pressure max.
bar at 0 liter
3.5
Notice:
If Galden
®
or Fluorinert
®
is used the the charge of the motor increases. Wrong
adjustment causes overheating and eventually destruction of the motor.
With the circulation pump P3 a maximum pump pressure stage >PUMP 1< may
be adjusted..
The pump type can be recognized in the order no. on the name plate.
95 xx xxx _ xx PX xx