Menu limits, Set min -94.99 – JULABO SC 2500w Professional Series Recirculating Coolers User Manual
Page 46

Menu functions
46
9.5. MENU
LIMITS
XXXXX
MENU
LIMITS
LIMITS: Press enter
to switch to menu level 2
>SET MAX<
>SET MIN<
>HEATMAX<
>COOLMAX<
>INT MAX<
>INT MIN<
>BAND H<
>BAND L<
In case of external control
these menu items are
additionally indicated.
1.
Press the
button until the desired submenu is displayed
2.
Set value:
Use the cursor keys to move left or right on the display until
the numeral you wish to change blinks.
Use the increase/decrease arrows
to change the selected
numeral (-, 0, 1, 2, 3, ... 9).
3.
Press enter
to store the value.
Continue: Press
/
or
Factory settings
XXXXX
SET MAX
300.00
XXXXX
SET MIN
-94.99
SETPOINT MAX / MIN -
Maximum and minimum setpoint
Restriction for the adjustable temperature range.
The limitation of the operating temperature range effects the temperature
setting under the menu called up with the
key.
It is possible to adjust only working temperatures that lie within the limit
range set here.
Existing settings for SETP 1, 2, 3 and also for >OVERTEMP< and
>SUBTEMP< (see page 29) are automatically defered within the limit
range.
Setting range:: -94.99 °C ... +300.0 °C