JULABO SC 2500w Professional Series Recirculating Coolers User Manual

Page 67

background image

Recirculating Cooler

67


Important times for a command transmission:

To ensure a safe data transfer, the time gap between two commands should
be at least 250 ms.
The circulator automatically responds to an in command with a data string
followed by a LF (Line Feed). The next command should only be sent after
10 ms.

The out commands are valid only in remote control mode.


Examples:
Command to set the working temperature >Setpoint1< to 55.5 °C

out_sp_00

55.5

Command to ask for the working temperature >Setpoint1<

in_sp_00

Response from the circulator:

55.5

LF



12.3. List of commands

out commands:

Setting temperature values or parameters.

Command

Parameter

Response of recirculating cooler

out_mode_01

0

Use working temperature > SETP 1<

out_mode_01

1

Use working temperature > SETP 2<

out_mode_01

2

Use working temperature > SETP 3<

out_mode_02

0

Selftuning „OFF“.
Temperature control by using the stored parameters.

out_mode_02 1

Selftuning

„ONCE“

Single selftuning of controlled system after the next start.

out_mode_02 2

Selftuning

„ALWAYS“

Continual selftuning of controlled system whenever a new
setpoint is to be reached.

out_mode_03

0

Set external programmer input to voltage.
Voltage

0 V ... 10 V

out_mode_03

1

Set external programmer input to current.
Current

0 mA ... 20 mA

out_mode_04

0

Temperature control of internal bath.

out_mode_04

1

External control with Pt100 sensor.

out_mode_05

0

Stop the unit = R –OFF-.