9.3.2. A-START
–
Autostart
XXXXX
A-START
OFF
Factory setting: OFF
The AUTOSTART function (automatic start mode) is allowing the start of
the instrument directly by pressing the mains power switch or using a
timer.
1.
Press the
button until the submenu > A-START < is displayed.
2.
Select the parameter with the keys
(OFF / ON).
3.
Press enter
to store the selected parameter.
Continue: Press
/
or
Possible parameters:
ON
- AUTOSTART on
OFF
- AUTOSTART off
Note:
The temperature system has been configured and supplied by JULABO
according to N.A.M.U.R. recommendations. This means for the start
mode, that the unit must enter a safe operating state after a power failure
(non-automatic start mode). This safe operating state is indicated by
OFF
, resp. R OFF on the VFD COMFORT-DISPLAY. A complete
shutdown of the main functional elements such as heater and circulating
pump is effected simultaneously.
The values set on the recirculating cooler remain stored, and the unit is
returned to operation by pressing the start/stop key (in manual control
mode).
In remote control mode, the values need to be resent by the PC via the
interface.
Should such a safety standard not be required, the AUTOSTART
function (automatic start mode) may be activated, thus allowing the start
of the instrument directly by pressing the mains power switch or using a
timer.
The AUTOSTART function can only be used, if setpoint setting is carried
out via > KEY <, > EPROG < or >PT100<.
Warning:
For supervised or unsupervised operation with the AUTOSTART function, avoid any
hazardous situation to persons or property.
The instrument does no longer conform to N.A.M.U.R. recommendations.
The safety and warning functions of the instrument should always be used to their
fullest capacity.