9.7. MENU ADJUST – ATC Absolute Temperature Calibration
ATC serves to compensate a temperature difference that might occur
between recirculating cooler and a defined measuring point in the bath
tank because of physical properties.
XXXXX
MENU
ADJUST
ADJUST: Press enter
to switch to menu level 2
>ATC SEN <
(INT)
or
(EXT)
> ATCSTAT <
> ATCSTAT <
(ON or OFF)
(ON or OFF)
>C ART
>C ART
>1- point <-, >2- point <- or >3- point < - calibration
> TTEMP 1<
> TTEMP 1<
> CTEMP 1<
> CTEMP 1<
> TTEMP 2<
> TTEMP 2<
> CTEMP 2<
> CTEMP 2<
> TTEMP 3<
> TTEMP 3<
> CTEMP 3<
> CTEMP 3<
1.
Press the
button until the desired submenu is displayed.
2.
Set value /parameter:
Use the cursor keys to move left or right on the display until
the numeral you wish to change blinks.
Use the increase/decrease arrows
to change the selected
numeral (-, 0, 1, 2, 3, ... 9).
or
select the parameter with the keys
.
3.
Press enter
to store the value / parameter
Continue: Press
/
or