background image

36

tina88e1-b (2014-09)

4.5 Switching

Functions

The two switching functions can be set to any pressure within
the measurement range of the gauge (

→  16).

The current setpoint setting
• is output at the analog port instead of the

measurement signal (

→  25) and can be measured with a

voltmeter after the button is pressed, or

• can be read/written via the RS232, Ethernet or wireless

interface.

• can be read/written via the diagnostic port,
If the pressure is lower than the setpoint, the corresponding LED
( or ) is lit solid and the corre-
sponding relay (

→  25) is energized.

Measurement signal (pressure p)

Time t

Me

asu

reme

nt value

Off

On

Hysteresis
(1% F.S.)

Off

Threshold value

Setpoint