Zeroing the Gauge

The gauge is factory calibrated while "standing upright"
(

→ "Calibration Test Report").

We recommend performing a zero adjustment, when the

gauge is operated for the first time.

Due to long time operation or contamination, a zero drift could
occur and zero adjustment may become necessary.
For adjusting the zero, operate the gauge under the same con-
stant ambient conditions and in the same mounting orientation
as normally.

4.4.1

<ZERO> Adjustment

The zero can be adjusted via

• the button on the gauge,
• the diagnostic port (→  [3]),
• briefly apply the supply voltage to the digital input

"Remote Zero" (pin 10),

• the RS232 interface (→  [2], [3]),
• the Ethernet interface (→  [1], [4]),
• the wireless network (→  [1], [4]).

While the gauge is being heated and/or under atmos-

pheric pressure, the zeroing function is locked in order
for operating errors to be prevented.