beautypg.com

1 diagnostic procedures, 1 measurement system diagnostic procedure – INFICON Q-pod Thin Film Deposition Monitor User Manual

Page 44

background image

4 - 16

IP

N 07

4-

54

7-

P1

B

Q-pod Operating Manual

4.3.1 Diagnostic Procedures

The following diagnostic procedures employ the Crystal Sensor Emulator to
analyze a constant Crystal Fail message. The symptom is a Crystal Fail message
that is displayed by the Q-pod software even after the monitor crystal has been
replaced with a new good monitor crystal.

4.3.1.1 Measurement System Diagnostic Procedure

1

Refer to

Figure 4-8 on page 4-15

. Remove the six-inch BNC cable from the

Feed-Through at point A.

2

Connect the Crystal Sensor Emulator to the 6 inch BNC cable at Point A.

If the Crystal Fail message disappears after approximately five seconds,
the measurement system is working properly. Re-install the six-inch BNC
cable to the Feed-Through. Go to

section 4.3.1.2

.

If the Crystal Fail message remains, continue at step 3.

3

Disconnect the six-inch BNC cable from the Oscillator and from the Emulator.

4

Visually inspect the six-inch BNC cable to verify that the center pins are seated
properly.

5

Use an Ohm meter to verify the electrical connections on the six-inch BNC
cable.

There must be continuity (<0.2 ohms, after accounting for resistance of
ohmeter leads) between the center pins.

There must be isolation (>10 megohms) between the center pins and the
connector shield.

There must be continuity between the connector shields.

Replace the six-inch BNC cable if it is found to be defective and repeat Step 2
of this procedure.

6

If the six-inch BNC cable is not defective, re-connect the six-inch cable to the
oscillator and to the Crystal Sensor Emulator. If the Crystal Fail message
remains, contact INFICON.