Vacuum settings
•
>
>
This menu allows to observe and to change the settings belonging to the vacuum
system.
6.6.1.1
Automatic purge (UL1000 Fab only)
•
>
>
>
Automatic purge (UL1000 Fab only)
Through this menu it is possible to program the automatic purge (
Please refer to
Chapter
) for 20 seconds when switching from measuring to standby mode.
6.6.1.2
Vent delay
•
>
>
>
Through this menu item it is possible to define the delay time until the inlet port is
vented when operating the STOP button. When the STOP button is pressed for a
period of time which is shorter than the delay time specified here, the UL1000 and
UL1000 Fab will just change to Stand-by mode.
When the STOP button is pressed for a period of time which is longer than the delay
time specified here, the UL1000 and UL1000 Fab will vent the inlet port.
Softkey 2:
Automatic purge (UL1000 Fab only)
Refer to chapter
Softkey 3:
Refer to chapter
Softkey 4:
Refer to chapter
Softkey 5:
Refer to chapter
Softkey 6:
Refer to chapter
Softkey 7:
Refer to chapter
Softkey 3:
OFF
Automatic purge is switched off at standby mode.
Softkey 7:
ON
Automatic purge is activated. When switching from measurement to STAND-BY the
forepump is rinsed automatically for 20 seconds.