Installation

31

iina

70

e 0

2

.f

m

te

chn

ica

l ha

nd

boo

k

(1

40

8)

2.3.2.5

RS232

This RS232 C interface

Fig. 2-6

/5 is wired as data communication equipment (DCE)

and permits the connection of a personal computer (PC) for monitoring and data
logging. The connection is made through a 9 pin sub-D socket. For more information
refer to the Interface Description.

2.3.2.6

Remote Control

This Remote Control interface

Fig. 2-6

/6 is a serial interface to control the UL1000

and UL1000 Fab by the RC1000. The RC1000 Remote Control can be connected via
the wireless transmitter or via an extension cable with a RJ45 plug. Refer to the
RC1000 Technical Handbook for more information.

Pin

Assignment

2

RXD

3

TXD

5

GND

7

RTS

8

CTS

Pin

Assignment

2

+24V (fuse 0.8 A time lag)

3

0 V

4

RXD (intern. RS232)

5

TXD (intern. RS232)

This manual is related to the following products: