Maintenance plan, 4 maintenance plan – INFICON UL1000 Helium leak detector User Manual
Page 101
Ma
intena
nce Work
101
iina70e 09.fm
technical handbook
(1408)
9.4
Maintenance Plan
Subassembly
Maintenance work
Operating hours/years
Service
level
Spare part
UL1000 / UL1000 Fab
1500
4000
8000
16000
24000
number
1/4
1
2
3
4
Vacuum system
Backing pump D16 B
Check oil level, if required exchange
X
I & II
Exchange the oil
2
X
II
20099091
Refurbish backing pump
X
III
Scroll pump Agilent
Exchange the tip seal
X
1
III
200001671
TS 620
Exchange the scroll head
X
III
200001665R
SplitFlow 80
Exchange the operating agent reservoir
X
2
II u. III
200003801
Bearing exchange
X
2
III
Valve block
Clean the valves, replace valve seals
2
X
III
200000594
Disassemble and clean the valve block
2
X
III
200000593
Replace filter of vent and purge gas line
1
X
1
I, II ,III
200000683
Align Pirani gauge
X
III
Silencer UL1000 Fab
Replace
X
1
I, II, III
20099183
Exhaust filter UL1000
Check, empty
X
I, II, III
Replace filter insert
X
1
I, II, III
200000694
Electrical system
Fan subassemblies
Blow out fan chassis wall and bottom
1
X
1
I
Replace replacement filter cell for chassis
wall fan
1
X
1
I
200000685