Maintenance Work

107

iina

70

e 0

9

.f

m

te

chn

ica

l ha

nd

boo

k

(1

40

8)

When inserting the new air filter insert, note the direction of the air flow. It is
indicated in the figure above

Fig. 9-2

by a black arrow.

Notice: The side of the filter insert marked „clean air side" or the white surface of

the filter insert must point in the direction of the fans.

Push the filter insert into the guide and place the covers back on. To close the
instrument, insert the covers and press these on.

Fig. 9-2 Replacement of the air filter in the chassis wall

Item

Description

Item

Description

a

Air filter cut out

2

Air filter guide

1

Air filter insert

3

Ejection bore (rear)

a

a

2

1

3