After power on a welcoming picture appears on the screen of the control panel

Fig.

3-1

/1, then status information on the speed of the turbo pump, the foreline pressure,

the emission and the active filament are given.

The start up procedure takes about 3 minutes and the end is indicated by a beep.
The UL1000 and UL1000 Fab is in Stand-by mode now.

4

Check if the inlet port

Fig. 3-1

/2 is blanked off. If not, please mount a blind flange with

O-Ring on the inlet port.

5

Press the

START Button

Fig. 3-2

/6. The inlet will be evacuated and the measured

leak rate will be displayed a moment later.

This is the measurement mode. If a test part was connected you would start spraying
Helium to identify leaks.