Error messages, Examples 1-4, Command list ascii mode – INFICON Helium Leak Detector Modul1000 Interface description User Manual
Page 6
1-4
ASCII Protocol
jin
s
80
e.f
m
Op
er
at
in
g M
a
nua
l
(1
01
1)
3.2
Error Messages
3.3
Examples
3.4
Command List ASCII Mode
OK
ERR_OK
command completed
E01
ERR_CMD_START
wrong command start (no "*")
E02
ERR_ERR_BLANK
illegal blank
E03
ERR_CMD_WORD_1
command word 1 illegal
E04
ERR_CMD_WORD_2
command word 2 illegal
E05
ERR_CMD_WORD_3
command word 3 illegal
E06
ERR_DISABLED
control by RS232 not enabled
E07
ERR_ARGUMENT
argument faulty
E08
ERR_NO_DATA
no data available
E09
ERR_BUFFER_OVERFLOW
error buffer overflow
E10
ERR_INVALID
command invalid
E11
ERR_NO_QUERY
query not allowed
E12
ERR_QUERY
only query allowed
E13
ERR_NOT_IMPLEMENTED
not yet implemented
Command
answer
*stat? (CR)
MEAS (CR)
mode
*status? (CR)
MEAS (CR)
mode
*read? (CR)
2.876E-7 (CR)
leak rate according to
programmed unit
*read:pa*m3/s? (CR)
2.876E-6 (CR)
leak rate in a different unit
*start (CR)
OK (CR)
start messurment
*conf:trig1? (CR)
1.0E-9
retrieve trigger 1
*conf:trig1 2.0E-9 (CR)
OK
set trigger 1
*CLS
clear error
*IDN
identification
:CRC
check sum
:DEVice
name of instrument (Modul1000)
:VERsion
software version
:SERial
serial-number
:TURBO
software version TC600
:MC68
hardware identification MC68
:IOversion
hardware identification IO panel
:GBversion
hardware identification control panel
:VDversion
hardware identification mother board
:DIP1
MC68 DipSwitch 1
:DIP2
MC68 DipSwitch 2