ASCII Protocol

jin

s

80

e.f

m

Op

er

at

in

g M

a

nua

l

(1

01

1)

3.2

Error Messages

3.3

Examples

3.4

Command List ASCII Mode

OK

ERR_OK

command completed

E01

ERR_CMD_START

wrong command start (no "*")

E02

ERR_ERR_BLANK

illegal blank

E03

ERR_CMD_WORD_1

command word 1 illegal

E04

ERR_CMD_WORD_2

command word 2 illegal

E05

ERR_CMD_WORD_3

command word 3 illegal

E06

ERR_DISABLED

control by RS232 not enabled

E07

ERR_ARGUMENT

argument faulty

E08

ERR_NO_DATA

no data available

E09

ERR_BUFFER_OVERFLOW

error buffer overflow

E10

ERR_INVALID

command invalid

E11

ERR_NO_QUERY

query not allowed

E12

ERR_QUERY

only query allowed

E13

ERR_NOT_IMPLEMENTED

not yet implemented

Command

answer

*stat? (CR)

MEAS (CR)

mode

*status? (CR)

MEAS (CR)

mode

*read? (CR)

2.876E-7 (CR)

leak rate according to
programmed unit

*read:pa*m3/s? (CR)

2.876E-6 (CR)

leak rate in a different unit

*start (CR)

OK (CR)

start messurment

*conf:trig1? (CR)

1.0E-9

retrieve trigger 1

*conf:trig1 2.0E-9 (CR)

OK

set trigger 1

*CLS

clear error

*IDN

identification

:CRC

check sum

:DEVice

name of instrument (Modul1000)

:VERsion

software version

:SERial

serial-number

:TURBO

software version TC600

:MC68

hardware identification MC68

:IOversion

hardware identification IO panel

:GBversion

hardware identification control panel

:VDversion

hardware identification mother board

:DIP1

MC68 DipSwitch 1

:DIP2