INFICON Helium Leak Detector Modul1000 Interface description User Manual
Page 34
1-32
List of menu parameter and ASCII/Binary protocol
jin
s
80
e.f
m
Op
er
at
in
g M
a
nua
l
(1
01
1)
Recorder output
scale Upper limit
Main Menu ->
Settings ->
Interfaces ->
Recorder ->
Recorder scale
1E-11mbarl/s to
1E7mbarl/s
1E-5mbarl/s
*CONFig:RECorder:
upperexp
Recorder scale
Main Menu ->
Settings ->
Interfaces ->
Recorder scale
0.5V/Decade to 10V/
Decade
1V/Decade
*CONFig:RECorder:
scale
Leak rate output
via recorder
output
Main Menu ->
Settings ->
Interfaces ->
Recorder scale
always, measurement
mode only
always
*CONFig:RECorder:
onlymeas
Date & Time
Main Menu ->
Settings ->
Miscellaneous ->
Time & Date
01.01.2005 12:00
*HOUR:DATE
*HOUR:TIME
Language
Main Menu ->
Settings ->
Miscellaneous ->
Language
ENGLISH, DEUTSCH,
ITALIANO, FRANCAIS,
POLSKI, NIHONGO
(KATAKANA), CHINESE,
ESPANOL
english
*CONFig:LANGuage
Leak rate filter
Main Menu ->
Settings ->
Miscellaneous ->
Leak rate filter
Fixed, I-CAL
I-CAL
*CONFig:ICAL
SetLrFilter
Part number
enable
Main Menu ->
Settings ->
Miscellaneous ->
Part number OR
Main Menu ->
Settings ->
Vacuum settings -
> Auto Leak Test
settings -> Part
number
0 bis 999999, disabled
disabled
*CONFig:PARTNO
SetPartNrEnabled
CAL settling time
Main Menu ->
Settings -> CAL
settling time
10s to 300s
10s
*CONFig:CALSETTL
INGTime
Service interval
air filter
Main Menu ->
Settings -> Service
interval air filter
500hrs. to 4000hrs.
1500Std.
*HOUR:SERVice:FIL
TER
Service message
air filter
Main Menu ->
Settings -> Service
message air filter
on,off
on
Calibration
request
Main Menu ->
Monitoring
functions ->
Calibration request
on,off
off
*CONFig:CALREQ
SetCalRequest
Menu
Parameter
Menu access
Range
Default factory
setting
ASCIIProtocol
BinaryProtokol