beautypg.com

INFICON Helium Leak Detector Modul1000 Interface description User Manual

Page 34

background image

1-32

List of menu parameter and ASCII/Binary protocol

jin

s

80

e.f

m

Op

er

at

in

g M

a

nua

l

(1

01

1)

Recorder output
scale Upper limit

Main Menu ->

Settings ->

Interfaces ->

Recorder ->

Recorder scale

1E-11mbarl/s to
1E7mbarl/s

1E-5mbarl/s

*CONFig:RECorder:
upperexp

Recorder scale

Main Menu ->
Settings ->
Interfaces ->
Recorder scale

0.5V/Decade to 10V/
Decade

1V/Decade

*CONFig:RECorder:
scale

Leak rate output
via recorder
output

Main Menu ->
Settings ->
Interfaces ->
Recorder scale

always, measurement
mode only

always

*CONFig:RECorder:
onlymeas

Date & Time

Main Menu ->
Settings ->
Miscellaneous ->
Time & Date

01.01.2005 12:00

*HOUR:DATE

*HOUR:TIME

Language

Main Menu ->
Settings ->
Miscellaneous ->
Language

ENGLISH, DEUTSCH,
ITALIANO, FRANCAIS,
POLSKI, NIHONGO
(KATAKANA), CHINESE,
ESPANOL

english

*CONFig:LANGuage

Leak rate filter

Main Menu ->
Settings ->
Miscellaneous ->
Leak rate filter

Fixed, I-CAL

I-CAL

*CONFig:ICAL

SetLrFilter

Part number
enable

Main Menu ->
Settings ->
Miscellaneous ->
Part number OR
Main Menu ->
Settings ->
Vacuum settings -
> Auto Leak Test
settings -> Part
number

0 bis 999999, disabled

disabled

*CONFig:PARTNO

SetPartNrEnabled

CAL settling time

Main Menu ->
Settings -> CAL
settling time

10s to 300s

10s

*CONFig:CALSETTL
INGTime

Service interval
air filter

Main Menu ->
Settings -> Service
interval air filter

500hrs. to 4000hrs.

1500Std.

*HOUR:SERVice:FIL
TER

Service message
air filter

Main Menu ->
Settings -> Service
message air filter

on,off

on

Calibration
request

Main Menu ->
Monitoring
functions ->
Calibration request

on,off

off

*CONFig:CALREQ

SetCalRequest

Menu
Parameter

Menu access

Range

Default factory
setting

ASCIIProtocol

BinaryProtokol