INFICON Helium Leak Detector Modul1000 Interface description User Manual
Page 20
1-18
Binary Protocol
jin
s
80
e.f
m
Op
er
at
in
g M
a
nua
l
(1
01
1)
18
SetAutomatik
Set automatic On/Off
No
uchar
0=Automatic off; 1=Automatic
On
20
21
GetTurboCtrl
SetTurboCtrl
Get / set TMP state
No
2 Bytes
Byte 0:
Nominal condition (0-off, 1-
reduced, 2-Normal)
Byte 1:
Actual condition (0-off, 1-
reduced, 2-normal, 3-up, 4-
down, 5-Err)
22
GetTurboF
Get TMP frequency
No
unsigned integer [Hz]
23
GetTurboI
Get TMP current
No
unsigned integer [0.01 A]
24
GetTurboP
Get TMP power
No
unsigned integer [W]
25
GetTurboH
Get TMP operation hours
No
unsigned long integer [h]
26
GetTCH
Get frequency converter operation hours
No
unsigned long integer [h]
27
GetTurboRunupTime
Get Turbo run-up time
No
unsigned integer [s]
28
29
GetExtValves
SetExtValves
Get / set external valves output
No
2 Bytes
Bit0 Valves Pin5 V30
Bit1 Valves Pin6 V31
Bit2 Valves Pin7 V32
Bit3 Valves Pin8 V33
Bit4 Valves Pin9 V34
Bit5 Valves Pin10 V35
Bit6 Valves Pin11 V36
Bit7 Valves Pin12 V37
Bit8 Valves Pin13 V20
Bit9 Valves Pin14 V21
Bit10 Valves Pin15 V22
The value 0xFFFF switches to
normal operatioin.
30
31
GetRangeVV
SetRangeVV
Get / set preamplifier range
No
unsigned char
0=13MOhm auto
1=470MOhm auto
2=15GOhm auto
3=-500GOhm auto
4=13MOhm manual
5=470MOhm manual
6=15GOhm manual
7=500GOhm manual
No
Name
Description
Parameter
Data format