INFICON Helium Leak Detector Modul1000 Interface description User Manual
Page 36
1-34
List of menu parameter and ASCII/Binary protocol
jin
s
80
e.f
m
Op
er
at
in
g M
a
nua
l
(1
01
1)
t_B zero delay
Main Menu ->
Settings ->
Vacuum settings -
> Commander
functions ->
Commander
timing -> t_B
Evacuation time
0.1s to 95s
30s
*CONFig:COMMAN
DTime:2
*CONFig:COMMAN
DTime:B
t_C charging time
Main Menu ->
Settings ->
Vacuum settings -
> Commander
functions ->
Commander
timing -> t_C
Charging time
0.1s to 95s
30s
*CONFig:COMMAN
DTime:3
*CONFig:COMMAN
DTime:C
t_D discharging
time
Main Menu ->
Settings ->
Vacuum settings -
> Commander
functions ->
Commander
timing -> t_D
Discharging time
0.1s to 95s
5s
*CONFig:COMMAN
DTime:4
*CONFig:COMMAN
DTime:D
t_E venting time
Main Menu ->
Settings ->
Vacuum settings -
> Commander
functions ->
Commander
timing -> t_E
Venting time
0.1s to 95s
30s
*CONFig:COMMAN
DTime:5
*CONFig:COMMAN
DTime:E
t_F ready to test
Main Menu ->
Settings ->
Vacuum settings -
> Commander
functions ->
Commander
timing -> t_F
Ready to test
0.1s to 95s
5s
*CONFig:COMMAN
DTime:6
*CONFig:COMMAN
DTime:F
t_G
measurement
period
Main Menu ->
Settings ->
Vacuum settings -
> Commander
functions ->
Commander
timing -> t_G
Measurement
period
0.1s to 95s
1s
*CONFig:COMMAN
DTime:7
*CONFig:COMMAN
DTime:G
Menu
Parameter
Menu access
Range
Default factory
setting
ASCIIProtocol
BinaryProtokol