Characteristic
external pressure
gauge
Main Menu ->
Settings ->
Interfaces ->
External pressure
gauge ->
Characteristic
linear 4 to 20mA, linear
0 to 10V, log 4 to 20mA,
log 0 to 10V
lin. 4-20mA
*CONFig:PEXT1:CH
ARacteristic
Zero point signal
external pressure
gauge
Main Menu ->
Settings ->
Interfaces ->
External pressure
gauge -> Zero
point
0mA to 20mA or
0V to 10V
4mA resp. 2V
*CONFig:PEXT1:ZE
ROU
Full scale signal
external presure
gauge
Main Menu ->
Settings ->
Interfaces ->
External pressure
gauge -> Full scale
0,1V to 10V or
0.2mA to 20mA
20mA resp. 10V
*CONFig:PEXT1:FU
LLU
Zero point
pressure external
pressure gauge
Main Menu ->
Settings ->
Interfaces ->
External pressure
gauge -> Zero
point
1E-11 mbar to
1E+04 mbar
1E0 mbar
*CONFig:PEXT1:ZE
ROP
Full scale
pressure external
pressure gauge
Main Menu ->
Settings ->
Interfaces ->
External pressure
gauge -> Full scale
1E-4 mbar to 5E5 mbar
1E4 mbar
*CONFig:PEXT1:FU
LLP
PLC output Pin
3…14
Main Menu ->
Settings ->
Interfaces ->
Define PLC
outputs
TRIGGER_1,
TRIGGER_2,
TRIGGER_3,
ZERO_ACTIVE,
EMISSION_ON,
MEASURE, STANDBY,
VENTED, ERROR,
WARNING,
CAL_ACTIVE,
CAL_REQUEST,
REC_STROBE,
GAS_BALLAST,
CYCLE_ACTIVE, SNIFF,
PUMP_DOWN, OPEN,
CLOSE
3 - TRIGGER_1
4 - TRIGGER_2
5 - TRIGGER_3
6 -
ZERO_ACTIVE
7 -
EMISSION_ON
8 - ERROR
9 -
CAL_ACTIVE
10 -
CAL_REQUEST
11 - OPEN
12 - OPEN
13 - OPEN
14 - OPEN
From
*CONFig:PLCOUTLI
NK:3to
*CONFig:PLCOUTLI
NK:14
Menu
Parameter
Menu access
Range
Default factory
setting
ASCIIProtocol
BinaryProtokol