INFICON Helium Leak Detector Modul1000 Interface description User Manual

Page 32

background image

1-30

List of menu parameter and ASCII/Binary protocol

jin

s

80

e.f

m

Op

er

at

in

g M

a

nua

l

(1

01

1)

Characteristic
external pressure
gauge

Main Menu ->
Settings ->
Interfaces ->
External pressure
gauge ->
Characteristic

linear 4 to 20mA, linear
0 to 10V, log 4 to 20mA,
log 0 to 10V

lin. 4-20mA

*CONFig:PEXT1:CH
ARacteristic

Zero point signal
external pressure
gauge

Main Menu ->
Settings ->
Interfaces ->
External pressure
gauge -> Zero
point

0mA to 20mA or
0V to 10V

4mA resp. 2V

*CONFig:PEXT1:ZE
ROU

Full scale signal
external presure
gauge

Main Menu ->
Settings ->
Interfaces ->
External pressure
gauge -> Full scale

0,1V to 10V or
0.2mA to 20mA

20mA resp. 10V

*CONFig:PEXT1:FU
LLU

Zero point
pressure external
pressure gauge

Main Menu ->
Settings ->
Interfaces ->
External pressure
gauge -> Zero
point

1E-11 mbar to
1E+04 mbar

1E0 mbar

*CONFig:PEXT1:ZE
ROP

Full scale
pressure external
pressure gauge

Main Menu ->
Settings ->
Interfaces ->
External pressure
gauge -> Full scale

1E-4 mbar to 5E5 mbar

1E4 mbar

*CONFig:PEXT1:FU
LLP

PLC output Pin
3…14

Main Menu ->
Settings ->
Interfaces ->
Define PLC
outputs

TRIGGER_1,

TRIGGER_2,

TRIGGER_3,

ZERO_ACTIVE,

EMISSION_ON,

MEASURE, STANDBY,

VENTED, ERROR,

WARNING,

CAL_ACTIVE,

CAL_REQUEST,

REC_STROBE,

GAS_BALLAST,

CYCLE_ACTIVE, SNIFF,

PUMP_DOWN, OPEN,

CLOSE

3 - TRIGGER_1

4 - TRIGGER_2

5 - TRIGGER_3

6 -
ZERO_ACTIVE

7 -
EMISSION_ON

8 - ERROR

9 -
CAL_ACTIVE

10 -
CAL_REQUEST

11 - OPEN

12 - OPEN

13 - OPEN

14 - OPEN

From
*CONFig:PLCOUTLI
NK:3to
*CONFig:PLCOUTLI
NK:14

Menu
Parameter

Menu access

Range

Default factory
setting

ASCIIProtocol

BinaryProtokol