Error messages, Command list, 4 error messages 4.5 command list – INFICON Helium Leak Detector Modul1000 Interface description User Manual
Page 18
1-16
Binary Protocol
jin
s
80
e.f
m
Op
er
at
in
g M
a
nua
l
(1
01
1)
4.4
Error Messages
4.5
Command List
230
RS232Host
command currently not allowed (host control)
231
RS232Fb
command currently not allowed (remote control)
232
RS232Invalid
command currently not allowed (i.e. CAL when
running up)
233
RS232PW1disable
pass word 1 disabled (menu function)
234
RS232PW2disable
pass word 2 disabled (service function)
235
RS232CmdFailed
execution of command failed
240
RS232Cmd
command does not exist
241
RS232FbChecksum
hand unit: checksum wrong
242
RS232FbTimeout
hand unit: timeout
243
RS232Len
number or length of part of parameter defective
244
RS232Para
parameter not in valid range
252
RS232Start
first sign wrong (unequal, 0x05)
253
RS232Checksum
transmitted and calculated checksum unequal
254
RS232Timeout
time out
255
RS232Buffer
buffer overflow
No
Name
Description
Parameter
Data format
0
SetASCII
Switch to ASCII protocol
1
GetP1
Get inlet pressure p1
Byte 0:
Unit (0-mbar, 1-Pa,
2-Torr)
float
2
GetP2
Get fore vacuum pressure p2
Byte 0:
Unit (0-mbar, 1-Pa,
2-Torr)
float
3
GetPext
Get pressure of external pressure gauge pext
Byte 0:
Unit (0-mbar, 1-Pa,
2-Torr)
float
5
GetDeviceID
Get device identification
No
uchar
(always 4 for Modul1000)
6
GetSWCheckSum
Get checksum leak detector software
No
usint
7
GetTurboVersion
Get TMP version number
No
string (6 Bytes)
8
GetVersion
Get software version (Main- and sub-version
number)
No
2 Bytes
Byte 0: Main version number
Byte 1: Sub version number