Trouble Shooting

jin

s

80

e.f

m

Op

er

at

in

g M

a

nua

l

(1

01

1)

5.2

ASCII Protocol

Error

Possible Reason

Solution

Modul1000 does not
reply / Modul1000 replies
after several command
with „E10“

„Carriage Return“ at the end of the
command is missing

Finish all commands with „Carriage Return“
(ASCII 0dhex / 13dez)

Modul1000 replies with
„E06“

Control via RS232 not enabled in the
Modul1000

Enable control via RS232 (see submenu

CONTROL

LOCATION

)

Modul1000 replies with
error message to the first
command only, following
commands are
interpreted correctly

Receiving buffer of the Modul1000
was not empty before sending the
first command (e.g. by plugging in the
RS232 cable during operation)

In the ASCII protocol the Modul1000 has not time out
function which will empty the receiving buffer
automatically. Therefore, the buffer should be emptied
before the first command by sending of ESC, ^C or ^X