INFICON Helium Leak Detector Modul1000 Interface description User Manual
Page 30
1-28
List of menu parameter and ASCII/Binary protocol
jin
s
80
e.f
m
Op
er
at
in
g M
a
nua
l
(1
01
1)
Beep sound
Main Menu ->
Trigger & Alarms -
> Volume
on,off
on
*CONFig:BEEP
SetBeepOnOff
Unit pressure
Main Menu ->
Trigger & Alarms -
> Units
mbar, Pa, Atm, Torr
mbar
*CONFig:UNIT:Pres
sure
SetUnit 0
Unit leak rate
Main Menu ->
Trigger & Alarms -
> Units
mbarl/s, Pam³/s, Atmcc/s,
Torrl/s, ppm, g/a,
oz/yr, atmcc/min
mbarl/s
*CONFig:UNIT:LR
SetUnit 1
Alarm delay
Main Menu ->
Trigger & Alarms -
> Alarm delay
0s to 600s
30s
*CONFig:ALARMDel
ay
Audio alarm type
Main Menu ->
Trigger & Alarms -
> Audio alarm type
Leckrate proportional,
Alarmtrigger, Setpoint,
Pinpoint
Trigger alarm
*CONFig:AUDio
SetTriggerAlarmTyp
Purge & Gas
ballast
Main Menu ->
Settings ->
Vacuum settings -
> Purge & Gas
ballast
Manual purge, Manual
Gas ballast, Automatic
purge
man. Gasballast
*CONFig:PURGe
Vent delay
Main Menu ->
Settings ->
Vacuum settings -
> Vent delay
immediately, 1s, 1.5s, 2s,
no vent
2s
*CONFig:VENTdelay
SetVentDelay
Partial flow
Evacuation
Main Menu ->
Settings ->
Vacuum settings -
> Partial flow
Forepump,
Partialflowpump, Fore-
and Partialflowpump
without partial
flow pump
*CONFig:PARTIALFl
ow:EVACuation
Partial flow
Measurement
mode
Main Menu ->
Settings ->
Vacuum settings -
> Partial flow
Forepump, Fore- and
Partialflowpump
without partial
flow pump
*CONFig:PARTIALFl
ow:MEASure
Measurement
period
Main Menu ->
Settings ->
Vacuum settings -
> Auto Leak Test
settings ->
Measurement
period
1s to 1800s
10s
*CONFig:TESTINGT
ime
Series error
message
Main Menu ->
Settings ->
Vacuum settings -
> Auto Leak Test
settings -> Series
error message
disable, 1 to 9
5
*CONFig:SERIESErr
msg
GetSeriesErrMsg
Part number
enable
on,off
off
*CONFig:PARTCOU
NT
SetPartNrEnabled
Menu
Parameter
Menu access
Range
Default factory
setting
ASCIIProtocol
BinaryProtokol