INFICON Helium Leak Detector Modul1000 Interface description User Manual
Page 25
Binary Protocol
1-23
jin
s
80
e.f
m
Op
er
at
in
g M
a
nua
l
(1
01
1)
102
GetUp
Get pressure gauge voltage
(internal & external)
Byte 0:
1=p1
2=p2
3=pext1
4=pext2
float [mV]
105
GetElTa
Get electronic temperature
No
float [°C]
106
GetEvsTa
Get preamplifier temperature
No
float [°C]
107
GetMiap
Get anode voltage
No
float [V]
108
GetMikp
Get cathode voltage
No
float [V]
109
GetMisp
Get suppressor voltage [V]
No
float [V]
110
GetMiakp
Get anode-cathode voltage
No
float [V]
111
GetUValve
Get valve voltage
No
float [V]
112
GetUf3f4
Monitoring voltage of fuse F3 and F4
No
float
113
GetUf1f2
Monitoring voltage of fuse F1 and F2
No
float
114
GetU15
Monitoring voltage of +/-15V
No
float
119
ParaReset
Reset parameter
Byte 0:
1=load factory
default settings
No
122
123
GetCtrlMode
SetCtrlMode
Get / set control location
No
uchar
0=local
1=RS232
2=PLC
3=RS232 & local
4=local & PLC
124
GetMeasRange
Get measure range
No
uchar
125
GetAutoLeakTestResu
lt
Get AutoLeakTest result
No
uchar
126
GetLrMaxV24
Get max. leak rate since last query over RS232
Byte 0:
Unit (0-mbar*l/s,
1-Pa*m³/s, ....)
float
127
GetTau
Get current equivalent tau (I*CAL only)
No
usint
128
GetDrift
Get current drift
No
float
129
GetCalMode
Get calibration mode
No
uchar
130
131
GetPartNrEnabled
SetPartNrEnabled
Get / set part number enabled
No
uchar
No
Name
Description
Parameter
Data format