SetSecInMeas
Time, since device is in measurement mode
No
unsigned integer [s]
10
11
GetValve
SetValve
Get / set Internal valves
For set commando (11) automatic must be
switched off (see command 18)
No
unsigned integer
V1 0x0001
V3 0x0002
V7 0x0004
V6 0x0008
V2 0x0010
V4 0x0020
12
13
GetValveCtrl
SetValveCtrl
Get / set lowered valve voltage
Byte 0:
0-lowered, 1- high,
2-auto
unsigned char
14
GetSpsInput
Get PLC input state
No
unsigned integer
Bit0 DigitalIn Pin3
Bit1 DigitalIn Pin4
Bit2 DigitalIn Pin5
Bit3 DigitalIn Pin6
Bit4 DigitalIn Pin7
Bit5 DigitalIn Pin8
Bit6 DigitalIn Pin9
Bit7 DigitalIn Pin10
Bit8 Accessories Pin 3
Bit9 Accessories Pin 6
15
SetRelais
Set digital output (only for test purposes)
No
unsigned integer
Bit0 DigitalOut Pin3
Bit1 DigitalOut Pin4
Bit2 DigitalOut Pin5
Bit3 DigitalOut Pin6
Bit4 DigitalOut Pin7
Bit5 DigitalOut Pin8
Bit6 DigitalOut Pin9
Bit7 DigitalOut Pin10
Bit8 DigitalOut Pin11
Bit9 DigitalOut Pin12
Bit10 DigitalOut Pin13
Bit11 DigitalOut Pin14
Bit12 Accessories Pin 4 & 7
Bit13 Accessories Pin 5 & 8
Values
≥ 65280 switch back to
normal mode.
17
SetRecorder
Set recorder output (only for test purposes)
Values greater than 10000 switch back to
normal mode.
Byte 0:
0=port 1, 1=port 2
unsigned integer
[mV]
No
Name
Description
Parameter
Data format