INFICON Helium Leak Detector Modul1000 Interface description User Manual
Page 31
List of menu parameter and ASCII/Binary protocol
1-29
jin
s
80
e.f
m
Op
er
at
in
g M
a
nua
l
(1
01
1)
Machine factor
Main Menu ->
Settings ->
Vacuum settings -
> Machine factor
1E-6 to 1E6
1,00E+00
*FACtor:MACHine
SetMFac
leak rate external
test leak for
sniffer mode
Entry only at start
of calibration
1E-6mbarl/s to
1E-mbarl/s
1E-5mbarl/s
*CONFig:CAL:EXTS
niff
SetTL 2
Leak rate internal
test leak
Main Menu ->
Settings ->
Vacuum settings -
> Leak rate
internal test leak
1E-9mbarl/s to
1E-5mbarl/s
1E-6mbarl/s
*CONFig:CAL:INT
SetTL 0
leak rate external
test leak for
vacuum mode
Entry only at start
of calibration
1E-9mbarl/s to
1E-3mbarl/s
1E-7mbarl/s
*CONFig:cal:EXTVa
c
SetTL 1
Background
suppression
Main Menu ->
Settings -> Zero &
Background ->
Background
suppression
off, inlet area,
internal only
intern only
*CONFig:SUPPRess
ion
SetSuppression
Vacuum
Backgroundsuppr
ession
Autoleaktest
Main Menu ->
Settings ->
Vacuum settings -
> Reference
measurement
off, on
on
*CONFig:SUPPRess
ion:AUTOLeaktest
SetSuppression
Auotleaktest
Backgroundsuppr
ession
Commander
Main Menu ->
Settings ->
Vacuum settings -
> Commander
functions ->
Backgrund
suppression
off, on, stable
on
*CONFig:SUPPRess
ion:COMMAnder
SetSupptession
Commander
Zero
Main Menu ->
Settings -> Zero &
Background ->
Zero
disable, enabled, I-Zero
enabled
*CONFig:ZERO
SetZeroMode
Mass
Main Menu ->
Settings -> Mass
2, 3, 4
4
*CONFig:MASS
Control location
Main Menu ->
Settings ->
Interfaces ->
Control location
PLC, RS232, local and
PLC, local and RS232,
local, all
local
*CONFig:CONTrol
SetCtrlMode
RS232 protocol
Main Menu ->
Settings ->
Interfaces ->
RS232 protocol
Binary, ASCII,
UL2xxLeakware
ASCII
*CONFig:RS232
SetASCII
Menu
Parameter
Menu access
Range
Default factory
setting
ASCIIProtocol
BinaryProtokol