27

22.

Press the following key sequence: -, -, +. The display now reads SCAL
450.

23.

Adjust the simulator to supply a 450

E F signal. Wait for 30 seconds while

the control stabilizes.

24.

Press the following key sequence: +, +, -. The display now reads SENS
450.

25.

To verify proper calibration, adjust the simulator to supply a 350

E F signal.

Within 30 seconds, the display should stabilize and read SENS 350.

26.

Press the Manual key. The display reads [MANUAL]. If the control did
not calibrate properly repeat steps 12 - 25.

27.

Press the Power on pushbutton to de-energize oven.

28.

Disconnect AC power to the oven.

29.

Remove Protocol controller to expose thermocouple inputs terminals.

30.

Disconnect the two pieces of type J thermocouple lead wire connected to
the Control T/C and Hi-limit T/C terminals.

31.

Re-connect control and Hi-Limit thermocouples to the controller
thermocouple terminals (Control T/C, and Hi-limit T/C).

32.

Re-install the Protocol controller onto the oven.

33.

Re-connect AC power to the oven.

34.

Press the Power on pushbutton to energize oven.

35.

Press the Tune key. The display reads [TUNE ].

36.

Reset any tune mode parameters that were changed in step 12 to perform
calibration (examples: DEG = F and CZO = 0.0).

37.

When changes have been completed, press the Manual key. The display
reads [MANUAL ].

The calibration procedure is complete.