26

15.

Apply a 250

E F signal to the high limit thermocouple input:

a.

Connect the piece of type J thermocouple lead wire, wired to the High
Limit T/C terminals, to a thermocouple simulator.

b.

Set the simulator to output a type J thermocouple signal.

c.

Twist together or jumper the piece of type J thermocouple lead wire, wired
to the Control T/C terminals. This creates a junction and prevents a
Control sensor error [S-T/C ERR] caused by an open thermocouple.
Press the Reset key to clear a Control sensor error [S-T/C ERR] caused
by an open thermocouple.

d.

Adjust the simulator to supply a 250

E F signal. Wait for 30 seconds while

the control stabilizes.

16.

Press the following key sequence: -, -, +. The display now reads HCAL
450.

17.

Adjust the simulator to supply a 450

E F signal. Wait for 30 seconds while

the control stabilizes.

18.

Press the following key sequence: +, +, -. The display now reads HIL 450.

19.

To verify proper calibration, adjust the simulator to supply a 350

E F signal.

Within 30 seconds, the display should stabilize and read HIL 350.

20.

To calibrate the control sensor, press the – key. The display reads
[SCAL 250].

21.

Apply a 250

E F signal to the control thermocouple input:

a.

Connect the piece of type J thermocouple lead wire, wired to the Sensor
T/C terminals, to a thermocouple simulator.

b.

The simulator should be set to output a type J thermocouple signal.

c.

Twist together or jumper the piece of type J thermocouple lead wire to the
Hi-Limit T/C terminals. This creates a junction and prevents a hi-limit
sensor error [H-T/C ERR] caused by an open thermocouple. Press the
Reset key to clear a hi-limit sensor error [H-T/C ERR] caused by an
open thermocouple.

d.

Adjust the simulator to supply a 250

E F signal. Wait for 30 seconds while

the control stabilizes.