1) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right and the ESC key multiple times to tab left until

the CAL menu is highlighted and displaying the submenu items.

2) Use the UP and DOWN arrow keys to scroll through the sub menu items

until VELCITY is highlighted.

3) Press the LEFT and RIGHT arrow keys to scroll the value. When the

correct velocity value is being displayed, proceed to step 8.

4) Alternatively, press the ENTER key to display the Digits Edit Box.

5) Press the UP and DOWN arrow keys to scroll the highlighted value.

6) Press the LEFT and RIGHT arrow keys to scroll the digit locations.

7) Repeat steps 5 & 6 until the velocity value is correctly displayed.

8) Press the OK key to set the velocity and return to the menu screen, or ESC

to cancel entering the velocity.

9) Finally, press the MEAS key to return to the measurement screen.

11.7 Zero

The DFX-7 has a feature to manually adjust the electronic zero. This feature used in
conjunction with manually adjusting the velocity, are the primary features necessary