DFX-7 Ultrasonic Flaw Detector
27
The Range is the distance from the delay that will be viewable on the screen. It can
be adjusted to any value greater than the Delay. The procedures to adjust the range
are outlined below:
Setting Range
1) Press the MEAS key once to activate measure menu items. Press the
MEAS key multiple times to move right and the ESC key multiple times to
move left, until the RANGE cell is highlighted.
2) Press the UP, DOWN, LEFT, and RIGHT arrow keys to scroll the highlighted
value.
3) Alternatively, press the ENTER key to display the Digits Edit Box.
4) Press the UP and DOWN arrow keys to scroll the highlighted value.
5) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
6) Repeat steps 4 & 5 until the RANGE value is correctly displayed.
7) Press the OK key to return to the measurement screen, or ESC to cancel
entering the RANGE.
8) Finally, press the MEAS key to return to the measurement screen.