DFX-7 Ultrasonic Flaw Detector

69

11) Press the OK key to set the % SCRN HEIGHT and return to the menu

screen, or ESC to cancel entering the % SCRN HEIGHT.

Activating Draw Mode

4) Use the UP and DOWN arrow keys to scroll through the sub menu items

until DRAW is highlighted.

5) Press the LEFT and RIGHT arrow keys to toggle the DRAW feature ON.

6) Proceed to the next section – Drawing the TCG curve.

Drawing the TCG

1) With all the parameters set in the previous section, press the MEAS key

once to return to the main measurement screen.

Note: The user should now see a highlighted cursor displayed in the waveform

area, starting at the left most side of the display area.