2) Couple the transducer to the first calibration standard, and peak up on the

reflector, adjusting the gain to keep the signal height within the boundaries

of the waveform display area, and at a suitable height to allow for

attenuation at greater depths in the overall range to be tested.

3) Press the

LEFT and RIGHT

arrow keys to move the cursor over the

waveform that will be used to set the REFERENCE GAIN and draw the

DGS curve.

Note: The DGS curve uses 16 points to plot the curve.

4) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right, and the ESC key multiple times to tab left,

until the DGS menu is highlighted and displaying the submenu items.

Alternatively, press the LEFT and RIGHT arrow keys multiple times using

the left button wheel.