Dakota Ultrasonics CMX DL User Manual

Page 80

background image

Dakota Ultrasonics

76

and place the transducer in steady contact with the coating and sample or

actual test material. Be sure that the reading is stable and the repeatability

indicator, in the top left corner of the display, is fully lit and stable. Press the

MENU key once to activate the menu items tab. Press the MENU key

multiple times to tab right and the ESC key multiple times to tab left until the

CAL menu is highlighted and displaying the submenu items.

3) Use the UP and DOWN arrow keys to scroll through the sub menu items

until COATING 1PT is highlighted.

4) Press the ENTER key to display the Digits Edit Box.

5) Press the UP and DOWN arrow keys to scroll the highlighted value.

6) Press the LEFT and RIGHT arrow keys to scroll the digit locations.

7) Repeat steps 5 & 6 until the known thickness value is correctly displayed.

8) Press the OK key to calculate the velocity and return to the menu screen, or

ESC to cancel the coating one point calibration.

9) Finally, press the MEAS key to return to the measurement screen and begin

taking readings.