Dakota Ultrasonics CMX DL User Manual

Page 25

background image

CMX

DL

High Performance Material & Coating Thickness Gauge

21

5) Repeat steps 3 & 4 until the B-START or B-DEPTH value is correctly

displayed.

6) Press the OK key to set the B-START or B-DEPTH value and return to the

DISP menu, or ESC to cancel entering the B-START or B-DEPTH value.

Note: the adjusted value will appear next to the B-START or B-DEPTH

menu labels.

7)

Finally, press the MEAS key to return to the measurement screen and begin

taking readings.

DIGITS

B-SCAN

In the upper left corner of each of the display photos above, is the repeatability
indicator. The repeatability indicator is represented by five vertical bars and
represents how repeatable the measurements are. In regular measurement mode,
the CMX

DL

makes 8 measurements a second. In scan mode, the CMX

DL

makes 200

measurements a second. If the coating mode option is activated, the CMX

DL

makes

3 measurements a second in regular measurement mode and 65 measurements a
second in scan mode. When the CMX

DL

is idle, only the left vertical bar will be

displayed. However, when the CMX

DL

is making a measurement, four or five of the

bars should be displayed on the repeatability indicator. If fewer than four bars are
showing, the CMX

DL

is having difficulty achieving a stable measurement and the

thickness value displayed is potentially unstable.