18 top & bottom end caps – Dakota Ultrasonics CMX DL User Manual
Page 34
30
3.18 Top & Bottom End Caps
The top & bottom end panels are where all connections are made to the CMX
DL
. The
diagram above shows the layout and description of the connectors:
Transducer Connectors
Refer to Diagram: The transducer connectors, and battery cover/probe zero disk are
located on the CMX
DL
’s top end cap. The transducer connectors are of type Lemo
“00”. Note: There is no polarity associated with connecting the transducer to the
CMX
DL
.
Probe Zero Disk & Battery Cover
Refer to Diagram: The Battery cover is the large round disk shown in the diagram.
Note: This same disk is also used as a probe zero disk. Simply remove the cover
when replacing the batteries (3 AA cells). When performing a manual probe zero
function, simply place the transducer on disk making firm contact. Important: Be
sure to follow the polarity labels located on the back label of the CMX
DL
. Note:
Rechargeable batteries can be used, however they must be recharged outside of the
unit in a stand alone battery charger.
RS-232 Connector
Refer to Diagram: The RS-232 connector, located on the bottom end cap, is a 2 pin
female Lemo connector. It is designed to connect directly from the CMX
DL
to a
standard AT serial port on a PC. The cable supplied with the CMX
DL
is a Lemo to 9
pin serial cable. Note: This connector is also used to upgrade the CMX
DL
with the
latest version of firmware.
USB Serial to USB Converter Cable
A converter cable can be attached to the 9 pin serial cable in needed (part no. N-402-
0510).