Performing an Auto Probe Zero (Off Block)

Coating Probe Identified Coating Probe Not Identified

1) Be sure all couplant has been removed from the face of the transducer.

2) Press the OK key to perform the automatic probe zero, or ESC key to

cancel the zero operation.

Coating Probe Identified Coating Probe Not Identified

3) The screens illustrated above will be briefly displayed followed by the main

measurement screen. The CMX

DL

is ready to be calibrated.