CiDRA GVF-100 Gas Volume Fraction Monitoring System User Manual

Page 72

background image

Copyright © 2006 CiDRA Corporation

Page 9-21

20639-01 Rev 03

Level 1

Level 2

Level 3

Range

Description

Sensor
Check

PASS or FAIL.
(Indicate which sensors
failed test)

Performs health check
on each sensor

4-20mA Test

Test 4-20mA outputs
from 4 to 20mA

Manual testing of 4-
20mA output #1 and 2

Autoset Gain

1.0, 4.65, 21.55, 98.65

Provides auto
adjustment of pre-
amplifier setting based
on the current process
operating condition

Check/Set
Gain

1.0, 4.65, 21.55, 98.65

Provides for manual
checking and setting of
pre-amplifier gain.
Autoset Gain can
override set point

Gain

Test Gain

PASS or FAIL

Test preamp to
determine if gain falls
within range of AGC
parameters

RAM Test

PASS or FAIL

System memory test

Self Test

DPRAM Test

PASS or FAIL

Dual port memory test

Keyboard
Test

Red LED illuminates
indicating key is
functional

Test keyboard operation

Clear History

Reset

Resets data history

Passkey

Param Edit

Algorithm

Rarely used diagnostic
feature. Contact CiDRA
Technical Support

Passkey

System
Monitor

Diagnostics

Monitor

Sensor
Monitor

Factory Technical
Support diagnostic
feature