Communications (C):

Indicates a connection (Ethernet) or activity (Serial). No received
serial messages for 10 seconds will clear the Serial indicator.

HART Activity (H):

Indicates a recent HART message processed by the transmitter. No
received HART messages for 10 seconds will clear this indicator.

Write Protect (W):

Indicates configuration changes cannot be made to the transmitter.
Write Protect can be turned on and off via the menu or HART. Default
is OFF.

Event Log Update (!):

A failure event has been saved in the Event Log, accessible from the
front panel menu. Blank means no change to event log since last
accessed.

Activity (/-\l):

Indicator will cycle during normal system operation

9.5.1.3 Display

Examples

The following figure shows the display when the system is collecting
data to calculate a measurement. The ‘------‘ indicates the system is
unable to make a measurement, or an internal parameter was
changed that has caused the system to re-initialize.

G

V

F

------ %

S

O

S

------- ft/s

GF INITIALIZE MODE

GVF

/

Figure 53

Initialization Mode

In the following display the system is making a measurement. The
GVF reading is 2.016 % Entrained Air and has a corresponding Speed
Of Sound of 315.7 ft/s. The quality metric applied to the measurement
(SQ) is .72.