2 Safety
9
Eppendorf Research® plus — Operating manual
NOTICE!
NOTICE! Carry-over, contamination and incorrect dispensing results due to the
incorrect use of pipette tips.
The pipette tips are for single use only. Prolonged use can have a negative impact on
dispensing tasks.
Use the pipette tips only once.
Do not autoclave the epT.I.P.S. Dualfilter.
NOTICE!
NOTICE! Incorrect dispensing volume for special liquids and from temperature
differences.
Solutions which differ greatly from water in terms of their physical data, or temperature
differences between the pipette, pipette tip and liquid can result in incorrect dispensing
volumes.
Avoid temperature differences between pipette, pipette tip and liquid.
Make sure that the temperature is constant, between 20 and 25°C and at ±0.5°C.
Check the dispensing volume and readjust the pipette in case of deviations.